Improvements in the electrochemical generation of arsine at gold electrodes as a sample introduction strategy for arsenic determination in well-water samples by atomic absorption spectrometry

Lenys Fernández*, Patricio Espinoza-Montero, José Alvarado, Hicher Zambrano, Patricio Carrera, Pablo Pozo, Antonio Díaz-Barrios

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

The development and evaluation of a cathode consisting of a gold electrode covered by a polyaniline film, PANI/Au modified electrode, for electrochemical generation of arsine is described. The efficiency of arsine production using the new electrode was ascertained by comparison with corresponding hydride generation atomic absorption spectroscopy determinations of arsenic (As) in aqueous standards. The PANI/Au modified electrodes provide better sensitivity, accuracy and precision than the pure Au cathodes. The PANI/Au cathodes were prepared by controlled electrodeposition cycles of polyaniline films on top of Au electrodes. Cathodes obtained after 10 electrodeposition cycles showed the best performance, while the ones obtained after five voltammetry cycles produced films too thin and those obtained after 20 voltammetry cycles produced films too thick for the purpose of this work. Under optimised conditions, the As limit of detection in aqueous solutions, according to the (3σ) criterion, was 2.48 µg L−1. For accuracy determination, the modified electrode was applied to quantification of As(III) in acidified aqueous solutions of the certified standard reference material, NIST SRM 1643d, and in well-water samples containing possible interfering ions. Application of the Student’s t-test showed no significant difference between the expected and obtained results of the NIST SRM standard analysis at the 95% confidence level. Values for six replicate determinations of As(III) in the well-water samples showed close agreement with values obtained by analyses using hydride generation atomic absorption. Recoveries were in the range of 95–105% at test for acceptable accuracy in the analysis of real samples.

Original languageEnglish
Pages (from-to)844-858
Number of pages15
JournalInternational Journal of Environmental Analytical Chemistry
Volume98
Issue number9
DOIs
StatePublished - 15 Jul 2018

Bibliographical note

Publisher Copyright:
© 2018, © 2018 Informa UK Limited, trading as Taylor & Francis Group.

Funding

The authors thank Proyecto Prometeo de la SENESCYT, República del Ecuador and Decanato de Investigación y Desarrollo-Universidad Simón Bolívar, Caracas-Venezuela, for financial support.

FundersFunder number
Caracas-Venezuela
Decanato de Investigación y Desarrollo-Universidad Simón Bolívar
Secretaría de Educación Superior, Ciencia, Tecnología e Innovación

    Keywords

    • Arsenic
    • electrochemical hydride generation and atomic absorption determination
    • gold-polyaniline electrode